‹ Bharat Singh Rawat · CV

Supersonic gas-curtain IPM: Elmer field + Geant4 Monte Carlo

An independent re-simulation of a published beam-profile monitor, built from open tools and compared with the original CST results.

The study

The supersonic gas-curtain ionisation profile monitor (SGC-IPM) images a proton beam without touching it: the beam crosses a thin, 45°-tilted sheet of argon, ionises a tiny fraction of it, and an electrode stack pulls the Ar⁺ ions onto a micro-channel plate (MCP), whose image is the beam profile. Its extraction system was designed in CST and published by Parambil et al. (Instruments 10 (2026) 4). I rebuilt it from the paper alone, with open-source codes, to check how well an independent model reproduces their predictions and to add the physics a field-and-tracking study leaves out.

  1. GeometryRepeller, ten field rings and the MCP, with positions and both voltage sets taken from the paper.
  2. Field (ElmerFEM)Axisymmetric electrostatic solution on a gmsh mesh, exported as a 0.1 mm field map of the whole chamber.
  3. Monte Carlo (Geant4 11.2)A real proton beam crosses the curtain; ionisation, delta electrons and ion birth come from Geant4's EM physics (option 4 + PAI). Ions and electrons are tracked through the Elmer field in Geant4.
  4. Gas motionEach ion starts with the jet velocity and thermal spread; background ions start at rest-gas thermal speeds.
  5. Background and lightResidual-gas N₂⁺ ions from the whole beam path, and gas-light photons emitted and transported to the MCP.
  6. Detector and statisticsMCP efficiency vs impact energy, 25 µm binning, Gaussian fits; bootstrap, Poisson and four-seed errors on every number.

PROTON BEAM

Voltage 1 case100 MeV, 600 nA
Voltage 2 case28 MeV, 70 nA
Beam radius R0.5, 2.5, 5, 10, 15 mm
Transverse profileGaussian, σ = 0.41 R, cut at R
Protons simulated90 000 / 32 000 per case

GAS CURTAIN (ARGON)

Density3.9 × 10¹⁶ m⁻³
Sheet1 mm thick, 60 × 40 mm, 45° to the beam
Jet speed587 m/s
Jet temperature5 K
W-value (eV per ion pair)26.4 eV

BACKGROUND GAS (N₂, ASSUMED)

Pressure1 × 10⁻⁸ mbar
Density2.4 × 10¹⁴ m⁻³
Temperature300 K
Region70 × 64 × 50 mm around the curtain
W-value34.8 eV

Gas densities are raised 10⁷× inside Geant4 so enough collisions happen to give good statistics; every count is weighted back by 10⁻⁷ (the density is still far too low for multiple collisions). Electron collisions are switched off after birth for the same reason. The background pressure, the MCP efficiency curve and the gas-light yields are estimates, not measurements of this chamber. This re-simulation is my own follow-up to the paper and is not part of the published work.

Extraction system
p (beam)Ar⁺N₂⁺ (background)e⁻γ (gas light)

Image on the MCP

Profiles at the MCP

Ion arrival positions, X (beam direction) and Y (gas-jet direction).

Potential map

Elmer solution in the plane through the axis (mirrored), equipotentials every 100 V; electrodes in grey.

Axial potential φ(z)

Potential along the extraction axis, repeller → curtain → MCP, for both voltage sets; dots mark the electrodes and their set voltages.

Ion energy at the MCP

Spread grows with beam size: ions from the low end of the tilted curtain fall further.

Field along the axis

Ez on the extraction axis, from the Elmer solution.

Position response

Centroid at the MCP when the proton beam is moved (Voltage 1, 2.5 mm beam).

Realistic counts

Geant4 11.2 (standard EM option 4 + PAI in the curtain, 15.76 eV threshold); curtain 3.9 × 10¹⁶ m⁻³, 1 mm at 45°.

MCP detection efficiency

Weight given to each ion by its impact energy (curve after Oberheide et al. 1997, approximate). Shaded: energies the ions actually arrive with.

Ionisation cross-section, p + Ar

Relativistic Bethe (used here) against the measured data in the paper's Fig. 4, and the Rudd model.

Elmer vs CST (paper)

Same geometry and voltages; plate heights read from the paper's Figure 2 (±1 mm). Widths on both sides include the MCP detection efficiency and the 25 µm channel binning, with Gaussian fits.

Simulation vs the measured MCP image (paper Fig. 6)

Measured: the phosphor-screen images of Fig. 6 (28 MeV, 70 nA; gas jet ON and OFF), converted from their published false colours back to intensity. Black: this simulation (Voltage 2, 2.5 mm beam), solid = gas-jet ions, dashed = background ions, placed so that the background lines coincide. Magenta: the measured gas-jet spot (fitted, FWHM ellipse) and the centre of the measured background line.